Fluke MET/CAL Procedure ============================================================================= INSTRUMENT: Fluke 8024B: (2 year) CAL VER /5700 DATE: 08-Sep-97 AUTHOR: Fluke Corporation REVISION: $Revision: 2.5 $ ADJUSTMENT THRESHOLD: 70% NUMBER OF TESTS: 56 NUMBER OF LINES: 202 CONFIGURATION: Fluke 5700A CONFIGURATION: Fluke 8842A ============================================================================= # # Source: # Fluke 8024B instruction manual (PN 616052 July 1981). # # Compatibility: # MET/CAL 4.0 or later # # Subprocedures: # None # # Required Files: # 57_8024.bmp # # System Specifications: # TUR calculation is based on specification interval of the accuracy file. # The default 5700A accuracy file contains 90 day specs. # # Fluke makes no warranty, expressed or implied, as to the fitness # or suitability of this procedure in the customer's application. # # The 90 day specifications of the 5700A are used in TUR computations. # STEP FSC RANGE NOMINAL TOLERANCE MOD1 MOD2 3 4 CON 1.001 ASK- P F W 1.002 ASK+ K 1.003 MMFC C2 1.004 HEAD EQUIPMENT SETUP 1.005 DISP [32] WARNING 1.005 DISP HIGH VOLTAGE is used or exposed during the performance 1.005 DISP of this calibration. DEATH ON CONTACT may result if 1.005 DISP personnel fail to observe safety precautions. 1.006 DISP Turn the UUT on. 1.007 DISP Ambient temperature: 21C - 25C. 1.007 DISP Relative humidity: less than 80%. 1.007 DISP Allow the UUT to stabilize. 1.008 HEAD {DC VOLTAGE TESTS} 1.009 PIC 57_8024 1.010 HEAD DC VOLTAGE TESTS: {200mV Range} 1.011 5700 200 190.0mV 0.3U 2W 2.001 5700 200 -190.0mV 0.3U 2W 3.001 HEAD DC VOLTAGE TESTS: {2V Range} 3.002 5700 2 0.000V 0.001U 2W 4.001 5700 2 1.900V 0.003U 2W 5.001 HEAD DC VOLTAGE TESTS: {20V Range} 5.002 5700 20 19.00V 0.03U 2W 6.001 HEAD DC VOLTAGE TESTS: {200V Range} 6.002 5700 200 190.0V 0.3U 2W 7.001 HEAD DC VOLTAGE TESTS: {1000V Range} 7.002 5700 1000 1000V 2U 2W 8.001 HEAD {AC VOLTAGE TESTS} 8.002 HEAD AC VOLTAGE TESTS: {200mV Range} 8.003 RNG 200mV AC 8.004 5700 0.0mV S 2W 8.005 MEMI Enter UUT reading in millivolts AC: 8.006 MEME 8.007 MEMC mV +0.2U 9.001 5700 200 190.0mV 1.6U 100H 2W 10.001 5700 200 190.0mV 3.2U 2kH 2W 11.001 5700 200 190.0mV 9.9U 5kH 2W 12.001 HEAD AC VOLTAGE TESTS: {2V Range} 12.002 5700 2 1.900V 0.099U 5kH 2W 13.001 5700 2 1.900V 0.032U 2kH 2W 14.001 5700 2 1.900V 0.016U 100H 2W 15.001 5700 2 0.190V 0.015U 5kH 2W 16.001 HEAD AC VOLTAGE TESTS: {20V Range} 16.002 5700 20 19.00V 0.16U 100H 2W 17.001 5700 20 19.00V 0.32U 2kH 2W 18.001 5700 20 19.00V 0.99U 5kH 2W 19.001 HEAD AC VOLTAGE TESTS: {200V Range} 19.002 5700 200 100.0V 1.8U 2kH 2W 20.001 5700 200 100.0V 1U 100H 2W 21.001 HEAD AC VOLTAGE TESTS: {750V Range} 21.002 5700 750 750V 10U 100H 2W 22.001 5700 750 750V 10U 1kH 2W 23.001 HEAD {DC CURRENT TESTS} 23.002 HEAD DC CURRENT TESTS: {2mA Range} 23.003 5700 2 1.900mA 0.015U 2W 24.001 HEAD DC CURRENT TESTS: {20mA Range} 24.002 5700 20 19.00mA 0.15U 2W 25.001 HEAD DC CURRENT TESTS: {200mA Range} 25.002 5700 200 190.0mA 1.5U 2W 26.001 5700 200 -190.0mA 1.5U 2W 27.001 HEAD DC CURRENT TESTS: {2000mA Range} 27.002 5700 2000 1900mA 15U 2W 28.001 HEAD {AC CURRENT TESTS} 28.002 HEAD AC CURRENT TESTS: {2mA Range} 28.003 RNG 2mA AC 28.004 5700 0.000mA S 2W 28.005 MEMI Enter UUT reading in milliamps AC: 28.006 MEME 28.007 MEMC mA +0.002U 29.001 5700 2 1.900mA 0.059U 400H 2W 30.001 5700 2 1.900mA 0.059U 100H 2W 31.001 HEAD AC CURRENT TESTS: {20mA Range} 31.002 5700 20 19.00mA 0.31U 100H 2W 32.001 5700 20 19.00mA 0.31U 1kH 2W 33.001 HEAD AC CURRENT TESTS: {200mA Range} 33.002 5700 200 190.0mA 3.1U 1kH 2W 34.001 5700 200 190.0mA 3.1U 100H 2W 35.001 HEAD AC CURRENT TESTS: {2000mA Range} 35.002 5700 2000 1900mA 31U 100H 2W 36.001 5700 2000 1900mA 31U 1kH 2W 37.001 HEAD {RESISTANCE TESTS} 37.002 HEAD RESISTANCE TESTS: {200 Ohm Range} 37.003 5700 200 100.0Z 0.5U CW 38.001 HEAD RESISTANCE TESTS: {2 kOhm Range} 38.002 5700 2 1.000kZ 0.002U CW 39.001 HEAD RESISTANCE TESTS: {20 kOhm Range} 39.002 5700 20 10.00kZ 0.02U CW 40.001 HEAD RESISTANCE TESTS: {200 kOhm Range} 40.002 5700 200 100.0kZ 0.2U 2W 41.001 HEAD RESISTANCE TESTS: {2000 kOhm Range} 41.002 5700 2000 1000kZ 3U 2W 42.001 HEAD RESISTANCE TESTS: {20 MOhm Range} 42.002 5700 20 10.00MZ 0.21U 2W 43.001 HEAD {CONDUCTANCE TEST} 43.002 HEAD CONDUCTANCE TEST: {200nS Range} 43.003 5700 200 100.0nY 3U 2W 44.001 HEAD {PEAK HOLD TEST} 44.002 RNG 2V AC 44.003 5700 1.900V 100H S 2W 44.004 DISP Push the Peak Hold switch ON. 44.005 MATH MEM1 = 1.900 44.006 MEMI Enter UUT reading in volts: 44.007 MEME 44.008 MEMC V 0.067U 45.001 5700 0.3mV 100H O S 2W 45.002 DISP Count the number of digits the UUT changes 45.002 DISP prior to the beeper sounding.[D10000][7] 45.003 EVAL Did the UUT change less than 10 counts? 46.001 DISP Push the Peak Hold switch OFF. 46.002 HEAD {CONTINUITY TEST AUDIBLE TONE} 46.003 RNG 2kZ 46.004 EVAL Is the " ^ " indicator ON? 47.001 5700 0Z S CW 47.002 EVAL Is the " ^ " indicator OFF and the " v " indicator ON? 48.001 DISP Depress the AC/DC switch to 48.001 DISP activate the audible tone. 48.002 5700 0Z S CW 48.003 EVAL Is there an audible tone? 49.001 HEAD {BT INDICATOR} 49.002 DISP Connect a variable DC power supply to the Battery 49.002 DISP Elminator Connector. 49.002 DISP Connect the 8842A INPUT to the output of the variable 49.002 DISP power supply. 49.003 RNG 200mV 49.004 5700 190.0mV S 2W 49.005 DISP Adjust the variable power supply until the BT indictor 49.005 DISP appears in the UUT display. 49.006 8842 7.0V 0.5U 2W 50.001 5700 190.0mV S 2W 50.002 MESS Decrease the output of the variable supply until the 50.002 MESS 8842A displays +6V. 50.003 8842 6.0V S 2W 50.004 5700 190.0mV 0.2U 2W 51.001 RNG 2kZ 51.002 5700 0Z S CW 51.003 DISP Adjust the variable power supply for an 8842A reading of 51.003 DISP +10V. 51.004 8842 10.0V S 2W 51.005 5700 1.000kZ 0.002U CW 52.001 DISP Adjust the variable power supply for an 8842A reading of 52.001 DISP +6V. 52.002 8842 6.0V S 2W 52.003 5700 1.000kZ 0.002U CW 53.001 HEAD {LEVEL DETECTOR TEST} 53.002 RNG 200kZ 53.003 DISP Remove all connections to the 8024B and insert 53.003 DISP a test lead set into the V Ohm S and Common input 53.003 DISP terminals. As quickly as possible, strike the 53.003 DISP tips of the test leads together. Verify that the 53.003 DISP " v " indicator blinks each time the probe tips 53.003 DISP touch. 53.004 EVAL Did the " v " indicator blink each time? 54.001 DISP Repeat the test holding the test leads together 54.001 DISP and momentarily separating them. Insure that the 54.001 DISP " ^ " indicator blinks each time the connection 54.001 DISP is removed. 54.002 EVAL Did the " ^ " indicator blink each time? 55.001 ASK- U 55.002 HEAD {TEMPERATURE} 55.003 DISP Depress the TEMP degC range switch and release the 55.003 DISP kOhm/nS switch on the UUT. 55.004 DISP Connect the John Fluke Thermocouple accessory to the 55.004 DISP mA/degC and COMMON Inputs. 55.005 DISP Wipe the tip of the thermocouple accessory clean and 55.005 DISP place the tip between your thumb and index finger until 55.005 DISP until the display readings stabilize. 55.006 EVAL Does the UUT read between 34degC and 39degC? 56.001 END